Transmille 8000 Series User Manual

Page 52

Advertising
background image

8000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Page 51

PRT Measurement : Multimeter Setup (Option)


Press

SHIFT

then

Ω 4WIRE to select PRT measurement.


Select PROBE CONFIGURATION SINGLE DUAL

Select LINEARISATION IEC 751 ITS-90

Select PRIMARY DISPLAY AVERAGE FRONT REAR

Select PRT Coefficients ITS-90 Front ITS-90 Rear IEC 751 Front IEC 751 Rear

ITS-90 Front

Coefficients

ITS-90 Rear

Coefficients

IEC 751 Front

Coefficients

IEC 751 Rear

Coefficients

R

0

100

R

0

100

R

0

100 R

0

100

a<0°C

0

a<0°C

0

A 3.908299e-3 A 3.908299e-3

b<0°C

0

b<0°C

0

B -5.774999e-7 B -5.774999e-7

a>0°C

0

a>0°C

0

C -4.183e-12 C -4.183e-12

b>0°C 0

b>0°C

0

OK

OK

OK

OK

Reset coefficients

Reset coefficients

Reset coefficients

Reset coefficients


OK

Saves coefficients as entered

Reset Coefficients

Reset coefficient data back to defaults

Additional Settings

Description

DIGITS

Set measurement resolution

FILTER TIME

Set FILTER TIME :

250ms to 64s

Dynamic Mode for Automatic filter (reading stability based)

ANALOGUE FILTER

Set ANALOGUE FILTER mode : ON / OFF

INPUT TERMINALS

Set INPUT TERMINALS to use : FRONT / REAR

RATIO

Set RATIO mode : ON / VALUE / OFF

READING MIN / MAX

Display READING MIN / MAX measurement data

TRIGGER

AUTO / MANUAL

GUARD TO LOW

Set GUARD TO LOW mode : FLOATING / SIGNAL LOW

INTERNAL

TEMPERATURE

Displays the internal temperature