Guide to setting up a prt probe – Transmille 8000 Series User Manual

Page 53

Advertising
background image

8000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Page 52

Guide to setting up a PRT probe


1. Use digital control dial to choose

Probe Configuration

Ä PRT Coefficients Ä Primary Display Ä PRT Coefficients


2. Select Linearisation / Coefficient location ITS-90 Front ITS-90 Rear IEC 751 Front IEC 751 Rear

3. Enter Coefficient data (examples shown below) :

ITS-90 Front

Coefficients

ITS-90 Rear

Coefficients

IEC 751 Front

Coefficients

IEC 751 Rear

Coefficients

R

0

100

R

0

100

R

0

100 R

0

100

a<0°C

0

a<0°C

0

A 3.908299e-3 A 3.908299e-3

b<0°C

0

b<0°C

0

B -5.774999e-7 B -5.774999e-7

a>0°C

0

a>0°C

0

C -4.183e-12 C -4.183e-12

b>0°C 0

b>0°C

0

OK

OK

OK OK

Reset Reset

Reset Reset


Coefficient data entry example : 3.908299e-3

3

SHIFT

9 0 8 2 9 9

SHIFT

SHIFT

MENU

SHIFT

MENU 3




4. Select PROBE CONFIGURATION

SINGLE

DUAL


5. Select PRIMARY DISPLAY AVERAGE FRONT REAR as required

DECIMAL

POINT

EXP

MINUS SYMBOL