Parts diagram, Restoring factory calibration, Troubleshooting – UEi Test Instruments DMG200 User Manual

Page 11: Low temperature operation

Advertising
background image

DMG200 User’s Guide

2

Parts Diagram

DMG200 User’s Guide

11

Restoring Factory Calibration

The original factory calibration can be restored at any time by the following procedure:

1.

Turn the DMG200 power off.

2.

Hold

while pressing

. As soon as the power turns on, release

and then press it

quickly at least three times. The CAL indicator should illuminate, and

should show on

the Main Numeric Display. If not, turn the power off again and repeat.

3.

Press

five times followed by

.

4.

The gauge will beep 5 times and return to normal operation. The DMG200 is now reset to the

original factory calibration.

Troubleshooting

Under certain conditions, the display may read

or

. Please use the table

below to determine and fix the problem:

Display

Mode

Possible Problem

Solution

Sensor Contaminated

Clean the Vacuum Sensor

Normal

Operation or

Run Mode

Ambient Temperature too

Low

Turn the DMG200 off, warm

the vacuum port with your

hand, turn the DMG200 back

on.

Calibration

Sensor Contaminated

Clean the Vacuum Sensor

and Restart Calibration

Gauged Warmed too

Quickly

Restart Calibration. Allow the

Gauge to Warm Slowly

Calibration

Gauge Disturbed During

Calibration

Restart Calibration. Leave

the Gauge Undisturbed

During Warming Phase.

Low Temperature Operation

The DMG200 can operate accurately at temperatures as low as 10°F (-12°C). While operating below

freezing (32°F/0°C), the display update rate will slow from 3.5 readings every second to one reading

every two seconds.

For satisfactory battery life at low temperatures, a 9V Lithium battery is recommended.