Calibration test, Calibration, Lcd display – UEi Test Instruments DMG200 User Manual

Page 3

Advertising
background image

DMG200 User’s Guide

10

Calibration Test

The DMG200 should rarely require recalibration, though it may be necessary to know that your gauge

is calibrated properly for full accuracy. The Calibration Test mode assures you that the DMG200 is

calibrated to factory specifications. Test the calibration as follows:

1.

Turn the DMG200 power off.

2.

Important: Expose the DMG200 to atmospheric pressure.

3.

Press and hold (do not release) the

key for about 5 seconds.

4.

The display will show

if the instrument is calibrated properly.

5.

The display will show

if the instrument requires calibration. Please see the

Calibration section below.

Calibration

If the Calibration Test indicates recalibration is required, the gauge may be recalibrated. Unlike any

other vacuum gauge, the DMG200 can be easily recalibrated to factory specifications without any

special equipment, with the following procedure:

1.

For best results, clean the sensor with alcohol prior to calibration. Ensure the sensor is

completely dry before proceeding.

2.

Turn the DMG200 power off. If necessary, install a fully charged battery into the gauge.

3.

Place the dust cap over the vacuum fitting.

4.

Hold

while pressing

. As soon as the power turns on, release

and then press it

quickly at least three times. The CAL indicator should illuminate, and

should show on

the Main Numeric Display. If not, turn the power off again and repeat.

5.

Place the DMG200 in a clear Ziploc (resealable zipper storage) bag, press out any extra air, and

seal.

6.

Place the bagged DMG200 into a freezer with a temperature of less than –5°C (23°F).

7.

Allow the DMG200 to cool to below -2°C (28.4°F). At this point, the alarm will sound and the
display will change to

.

8.

Remove from freezer and press any key to silence the alarm.

9.

Place the DMG200 undisturbed in an area with a room temperature of at least 23°C (73.4°F) but

no greater than 30°C (86°F).

10. Allow the DMG200 to warm to 20°C (68°F). At this point, the alarm will sound, and

will show on the display.

11. Press any key to silence the alarm. The DMG200 is now calibrated to factory specifications.

Note: For accurate calibration, it is necessary to allow the DMG200 to warm slowly. Attempting to

accelerate the warming by using a heat source will not provide satisfactory results. During the

cooling/warming process, the temperature will be indicated on the Alternate Numeric Display in

degrees Celsius. The calibration process may be canceled at any time by turning off the DMG200 or

by pressing and holding

. The previous calibration will be unchanged.

DMG200 User’s Guide

3

LCD Display

Item

Function

A

Main Numeric Display

B

Alternate Numeric Display

C

“Analog” Vacuum Level Bar Graph
Backlight Indicator. Flashing: Temporary, Solid: Always On
Sound Indicator
Battery Level Indicator
Set Mode Indicator
Run Mode Indicator
Run Mode Complete Indicator
Oil Sensor Indicator
Calibration Mode Indicator
Indicates that Time is Displayed on the Alternate Numeric Display
Indicates that Temperature is Displayed on the Alternate Numeric Display
Indicates that the Vacuum Leak Rate is Displayed on the Alternate Numeric

Display
Indicates either Celsius or Fahrenheit Degrees are Displayed
Indicates that the Vacuum Leak Rate is Displayed in UNITS/second.
Pascal Units Indicator
Millibar Units Indicator
Millitorr Units Indicator
Millimeters of Mercury Units Indicator
Micron Units Indicator

Advertising