Meter problem solving – UEi Test Instruments SMART BELL PLUS KIT User Manual

Page 14

Advertising
background image

METER PROBLEM SOLVING

If any problems are not solved with these solutions, contact us or an authorized repair

center.

Fault symptom

Causes / Solutions

CO or CO2 too low

• Air leaking into probe, tubing, water trap

or connectors.

CO2 reading (- - - -)

• Meter was stored in a cold environment and

is not at normal working temperature.

Batteries not holding charge

• Batteries exhausted.

Meter does not respond to flue gas

• Particle filter blocked.

• Probe or tubing blocked.

• Pump not working or damaged with

contaminants.

Net temperature or Efficiency

• Ambient temperature set wrong during

calculation incorrect.

Automatic Calibration.

Flue temperature readings erratic

• Temperature plug reversed in socket.

• Faulty connection or break in cable or plug.

T flue displays (- - - -)

• Probe not connected or faulty

X-Air, EFF display

• CO2 reading is too low.

(- - - -)