AirSep MN165-1 B User Manual
Page 14
 
11
MN156-1 rev B
08/11
7.0 Troubleshooting
If your concentrator fails to operate properly, consult your Equipment Provider, and refer 
to the troubleshooting chart on the following pages for probable causes and solutions. 
Do not attempt any maintenance other than the possible solutions listed 
below. 
If the unit has not been used for an extended time period, it needs to 
operate for several minutes before power failure alarm can become 
activated. 
Problem
Probable Cause
Solution
Unit does not operate. Power 
failure condition causes an 
alarm to sound. 
The power cord is not 
connected to the 
electrical outlet. 
No power at the 
electrical outlet. 
 
Circuit breaker on 
concentrator unit is 
activated. 
Check power cord at 
the electrical outlet for a 
proper connection. 
Check power source, 
wall switch, in-house 
fuse or circuit breaker. 
Press (do not hold in) 
the circuit breaker reset 
button in the front of the 
unit. 
If the circuit breaker 
trips again or the alarm 
continues to sound after 
the unit is turned on, 
contact your Equipment 
Provider. 
Limited or no oxygen flow.
Dirty or obstructed 
humidifier bottle, or leak 
present. 
 
 
Defective nasal cannula, 
or other oxygen 
accessories, i.e. oxygen 
tubing. 
Cannula tubing loose.
 
 
Other leak or restriction.
Remove the humidifier 
bottle, and if flow is 
restored, clean or 
replace the humidifier 
bottle. 
Remove and check 
accessories for kinks or 
obstructions. Replace if 
needed. 
Check cannula tubing 
connection at control 
panel. 
Contact your Equipment 
Provider