AirSep MN165-1 B User Manual
Page 15
 
12
MN156-1 rev B
08/11
Problem
Probable Cause
Solution
Condensation collects in the 
oxygen tubing when you use 
a humidifier bottle 
 
 
Unit not properly 
ventilated. 
Elevated operating 
temperatures. 
Make sure unit is 
positioned away from 
curtains or drapes, hot 
air registers or heaters, 
Be certain to place the 
unit so all sides are at 
least 1 foot (30 cm) 
away from a wall or 
other obstruction. Do 
not place the unit in a 
confined area. 
Allow oxygen tubing to 
dry out, or replace with 
new tubing. Refill 
humidifier bottle with 
COLD water. 
DO NOT OVERFILL. 
Intermittent alarm sounds at 
one-second intervals. 
Equipment malfunction.
Set I/0 power switch to 
0 position, use your 
reserve oxygen supply 
(if provided), and 
consult your Equipment 
Provider immediately. 
All other problems.
Set I/0 power switch to 
the 0 position, use your 
reserve oxygen supply 
(if provided), and 
consult your Equipment 
Provider immediately.