AirSep Corp. — FreeStyle Service Manual

5-11

4. If either of the two fittings connecting the compressor to the manifold is connected

to the compressor, remove it and insert it into the manifold.
(See Figure 5.18.)

Figure 5.18

5. Connect the new compressor to the PSA assembly, and tighten the bolt to a torque of

15 in- lbs. (See Figure 5.19.)

Figure 5.19