Problem, Possible cause, Solution – AirSep MN126-1 C User Manual
Page 63
 
Problem
Possible Cause
Solution
Oxygen concentration is 
below specifications in 
test mode.
Unit’s temperature is too high. 
Blocked air intake or outlet; or 
defective fan. 
Leaks.
Restriction exhaust filter.
Defective 
Compressor 
Contaminated adsorbent beds.
Defective solenoid valve.
Clear any obstruction in
air intake or outlet filter; or
replace fan as needed.
Correct any leaks.
Replace exhaust filter.
Replace Compressor
Replace adsorbent beds.
Replace solenoid
valve.
Unit does not deliver
oxygen.
Cannula kinked or not properly
connected.
FreeStyle is not turned on.
Battery is dead.
Defective Universal AC / 
DC power supply.
OCD sensor defective.
OCD solenoid valve defective.
Check cannula and the
cannula connection.
Turn FreeStyle on to the
prescribed setting.
Replace battery.
Replace Universal AC / 
DC power supply.
Replace main circuit 
board.
Replace defective
OCD solenoid valve.
Unit continuously flows 
oxygen (instead of pulse flow 
delivery).
Service LED is on.
Service LED does not light 
at required 5,000-hour 
service. 
No hour meter display. 
 
Defective OCD valve. 
.
 
 
Service check required.
Defective keypad
 
 
 
Defective hour meter display.
Replace defective OCD 
solenoid valve.
Perform routine service 
and reset hour meter in 
maintenance mode.
Replace keypad
 
 
Replace hour meter. 
 
7-4
AirSep Corp. — FreeStyle Service Manual