20 — 20626704 C (English)

Optional Humidifier

If used, clean your humidifier daily to reduce the possibility of

contamination. Follow the cleaning recommendations provided

by the home care provider, or the manufacturer of the humidifier.

Replace the humidifier monthly, or as recommended by the

manufacturer, or your home care provider. (Do not overfill the

humidifier.)

WARNING: DO NOT USE LIqUID DIRECTLY ON THE UNIT.
A LIST OF UNDESIRABLE CHEMICAL AGENTS INCLUDES
BUT IS NOT LIMITED TO THE FOLLOWING: ALCOHOL AND

ALCOHOL-BASED PRODUCTS, CONCENTRATED CHLORINE-BASED
PRODUCTS (ETHYLENE CHLORIDE), AND OIL- BASED PRODUCTS
(PINE-SOL®, LESTOIL®). THESE ARE NOT TO BE USED TO CLEAN THE
PLASTIC HOUSING ON OXYGEN CONCENTRATOR, AS THEY CAN DAM-
AGE THE UNIT’S PLASTIC.

CAUTION: DO NOT operate the Eclipse and ambulate while
a humidifier is attached. Remove the humidifier bottle be-
fore walking. DO NOT lay the Eclipse down while attached

to a humidifier bottle.

Universal Cart

Before cleaning the Universal Cart, remove it from the Eclipse.

Use a damp (not wet) cloth or sponge and mild detergent solution

to clean the Universal Cart.

Power Cartridge (Battery)

The Battery in the Eclipse requires special care to assure a longer

life and the highest level of performance. The SeQual Battery is

the only approved Battery recommended for use with the Eclipse.

Use a damp (not wet) cloth or sponge to clean the Battery. First

spray the cloth or sponge with a mild detergent then clean the

Battery case and the latch.

SeQual Recommended Monthly Battery Service: Run the Eclipse

on Battery completely down to power loss condition (red light

and alarm). Plug Eclipse into AC power and recharge until full

on indicator. The device does not have to be running to charge

(which allows for shorter charge time).

SeQual recommends replacing the power cartridge after 2 years.

Battery Disposal

Your Battery is rechargeable and can be recycled. Always return

it to your home care provider for proper disposal. You can also

contact your local city or town offices for instructions on proper

disposal of the Battery.

AC Power Supply, DC Power Supply, Power Cord and Outer
Case:

Turn OFF the Eclipse and disconnect from AC or DC power

before any cleaning or disinfection activity. DO NOT spray the

outer case directly. Use a damp (not wet) cloth or sponge. Spray

the cloth or sponge with a mild detergent solution to clean the

cabinet and power supplies. To disinfect the Eclipse, use Lysol®

Brand II disinfectant (or equivalent). Proceed as directed by the

cleaner manufacturer.

WARNING: ELECTRICAL SHOCK HAZARD. TURN OFF THE
UNIT AND DISCONNECT THE POWER CORD FROM THE
ELECTRICAL OUTLET BEFORE YOU CLEAN THE UNIT TO

PREvENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD. ONLY
YOUR EqUIPMENT PROvIDER OR A qUALIFIED SERvICE TECHNICIAN
SHOULD REMOvE THE COvERS OR SERvICE THE UNIT. DO NOT USE
LIqUID DIRECTLY ON THE UNIT. A LIST OF UNDESIRABLE CHEMICAL
AGENTS INCLUDES BUT IS NOT LIMITED TO THE FOLLOWING: AL-
COHOL AND ALCOHOL-BASED PRODUCTS, CONCENTRATED CHLO-
RINE-BASED PRODUCTS (ETHYLENE CHLORIDE), AND OIL- BASED
PRODUCTS (PINE-SOL®, LESTOIL®). THESE ARE NOT TO BE USED TO
CLEAN THE PLASTIC HOUSING ON OXYGEN CONCENTRATOR, AS THEY
CAN DAMAGE THE UNIT’S PLASTIC.
WARNING: ELECTRICAL SHOCK HAZARD. DISCONNECT THE POWER
CORD FROM THE ELECTRICAL OUTLET BEFORE YOU CLEAN THE UNIT
TO PREvENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD.
ONLY YOUR EqUIPMENT PROvIDER OR A qUALIFIED SERvICE TECHNI-
CIAN SHOULD REMOvE THE COvERS OR SERvICE THE UNIT.
WARNING: CARE SHOULD BE TAKEN TO PREvENT THE OXYGEN CON-
CENTRATOR AND THE BATTERY FROM GETTING WET OR ALLOWING
FLUIDS TO ENTER THE UNIT. THIS CAN CAUSE A MALFUNCTION OR
SHUT DOWN, AND CAUSE AN INCREASED RISK FOR ELECTRICAL
SHOCK OR BURNS.
WARNING: CLEAN THE CABINET, CONTROL PANEL, AND POWER
CORD ONLY WITH A MILD HOUSEHOLD CLEANER APPLIED WITH A
DAMP (NOT WET) CLOTH OR SPONGE, AND THEN WIPE ALL SURFACES
DRY. DO NOT ALLOW ANY LIqUID TO GET INSIDE THE DEvICE. PAY
SPECIAL ATTENTION TO THE OXYGEN OUTLET FOR THE CANNULA
CONNECTION TO MAKE SURE IT REMAINS FREE OF DUST, WATER,
AND PARTICLES.