English – Merit Medical ASAP Aspiration Catheter User Manual

Page 2

Advertising
background image

Aspiration Catheter

I N S T R u C T I O N S f O R u S E

INTENdEd uSE

The Merit ASAP Aspiration Catheter is intended for the removal of fresh, soft emboli and

thrombi from vessels of the arterial system.

CAuTIONS

RX Only: Caution: Federal Law (USA) restricts this device to sale by or on the order

of a physician.

Read instructions prior to use

Store in a cool, dry place

Inspect contents prior to use

Do not expose to organic solvents such as alcohol.

Product is intended for single use only.

Do not reuse or re-sterilize; do not autoclave.

Do not use if package is opened or damaged.

The ASAP Aspiration Catheter should be used by physicians with adequate training in the use

of the device.

Kit components should not be substituted.

Contents of unopened, undamaged package are sterile and non-pyrogenic.

Crossing a freshly deployed drug eluting stent could damage the delicate drug coating.

wARNINGS

Do not use a bent, kinked or damaged catheter as this may lead to vessel injury and/or an

inability to advance or withdraw the catheter.

Do not advance the guide wire if resistance is met.

Do not place more than 60mls of fluid in the MicroStop fluid collection basin.

The ASAP Aspiration catheter must be used with a guide catheter with a minimum internal

diameter of 0.070”/1.78mm.

Do NOT flush the system while the catheter is still inside the patient vasculature.

POTENTIAl COMPlICATIONS

Potential complications include, but are not limited to:

Local or systemic infection; local hematomas; intimal disruption; arterial dissection; perfora-

tion and vessel rupture; arterial thrombosis; distal embolization of blood clots and plaque;

arterial spasm; arteriovenous fistula formation; catheter fracture with tip separation and

distal embolization; acute myocardial infarction; emergent surgery; death; abrupt closure of

total occlusion of treated graft or vessel; distal embolization of debris resulting in pulmonary

compromise and/or limb ischemia; death, myocardial infarction; coronary or bypass graft

thrombosis or occlusion, myocardial ischemia; stroke/CVA; emergent or non-emergent

fibrillation; hemorrhage; hypotension; pseudo aneurysm at access site.

Risks normally associated with percutaneous diagnostic and/or interventional procedures.

CONTRAINdICATIONS

Do not use in vessels less than 2.0mm in diameter.

The venous system

The removal of fibrous, adherent or calcified material (i.e. chronic clot, atherosclerotic plaque)

Not for use in the cerebral vasculature.

AddITIONAl EquIPMENT REquIREd buT NOT SuPPlIEd

Guiding catheter with an internal diameter of at least 0.070”/1.78mm

Guide wire with diameter of ≤0.014”/0.36mm

Rotating hemostasis valve

Sterile, heparinized normal saline for system flushing

PROduCT dESCRIPTION

The ASAP Aspiration Catheter Kit contains a dual lumen rapid exchange catheter, compatible

with 0.014”/0.36mm guide wires with related accessories. The catheter has a maximum outer

diameter of 0.068”/1.73mm and a working length of 140cm. The catheter is placed through a

6F guide catheter with a minimum inner diameter of 0.070”/1.78mm. The catheter has a radi-

opaque marker band located approximately 2mm proximal to the distal tip. The catheter has

three (3) non-radiopaque positioning marks located approximately 90cm, 100cm, and 110cm

proximal of the distal tip. Catheter has a distal hydrophilic coated section of 20cm.

The kit consists of the following components. These components may be packaged together

or separately.

(1) ASAP Aspiration Catheter

(2) 30ml VacLok® syringes

(2) 70 micron pore filter baskets

(1) MicroStop fluid collection basin

(1) Extension tubing set (8 ½”/21.5cm total length) with one-way stopcock

(1) RXP™ flush syringe (4ml)

INSTRuCTIONS fOR uSE

The following instructions provide technical direction but do not obviate the necessity of for-

mal training in the use of extraction or aspiration catheters. The techniques and procedures

described do not represent all medically acceptable protocols, nor are they intended as a

substitute for the physician’s experience and judgment in treating any specific patient.

Preparation and Directions for Use:

1. Open the ASAP Aspiration Kit box. Using aseptic technique, open the pouch and transfer

the tray onto the sterile field.

2. Remove the catheter hoop containing the ASAP catheter and other kit components from

the tray. Attach a 10ml syringe, (or one of the 30ml VacLok syringes included in

kit), filled with heparinized saline to the flush port on catheter hoop; flush hoop

completely to activate hydrophilic coating on distal portion of catheter.

3. Inspect the catheter for any bends or kinks. Remove the wire stylet from the rapid

exchange lumen.

4. Use 4ml RXP syringe filled with heparinized saline to flush rapid exchange lumen.

5. Flush extension tubing set with heparinized saline prior to use. Using one of the VacLok

syringes, draw 5-10ml of heparinized saline into the syringe. Attach the extension tubing

set with stopcock to the catheter. Turn the stopcock to the “open” position and flush with

the heparinized saline solution to remove all air from the system.

6. Turn the stopcock to the “off/closed” position after the catheter system is properly flushed.

Verify that the stopcock on the extension tubing set is in the closed position and connect

the VacLok syringe to the tubing set. Check that all fittings are secure so that air is not

introduced into the extension line during aspiration/extraction.

7. VacLok syringe setup: To create and maintain vacuum, withdraw the VacLok syringe

plunger to the desired position and rotate the plunger clockwise to position one of the

locking fins behind the stop pin. Turn the plunger counterclockwise to release. (see fig. A)

8. Check that all fittings are secure so that air is not introduced into the extension

line or syringe during aspiration. At this time verify that the extension tubing

set with stopcock is on the catheter (stopcock in the “closed” position) with

VacLok syringe attached . The ASAP catheter is completely prepped and is

ready for use.

Figure A

uSE Of THE ASAP CATHETER duRING AN INTERvENTIONAl PROCEduRE

Perform aspiration using the ASAP Catheter:

9. Load the prepped ASAP Catheter over the ≤0.014”/0.36mm guide wire

10. Confirm that the tubing set/VacLok have been connected to the catheter prior to

placement through guide catheter.

11. Advance the ASAP Catheter through the guide catheter, under fluoroscopy, and continue

to advance the catheter over the guidewire to the selected vascular site. The ASAP

catheter has three non-radiopaque positioning marks indicating 90cm, 100cm & 110cm

from the catheter tip. Position the distal tip marker proximal to the desired site. Stop

advancement of the ASAP Catheter if any resistance is encountered.

wARNING: Never advance or withdraw an intravascular device against resistance until the

cause of the resistance is determined by fluoroscopy. Movement of the catheter or guide wire

against resistance may result in damage to the catheter, or vessel perforation.

12. After fluoroscopically confirming catheter position, open the stopcock to begin aspiration.

Advance the catheter slowly distally away from the guiding catheter. Blood will enter the

VacLok Aspiration Syringe until the entire vacuum is gone (or the VacLok Aspiration

Syringe is filled).

a. If the syringe does not begin to fill with blood within 5 seconds, close stopcock and

remove ASAP catheter. Flush the catheter (extraction lumen) or use a new catheter.

Repeat steps 9-11 .

warning – Do not flush the system while the catheter is still inside the patient vasculature.

b. After completing the aspiration process, turn the stopcock to the “Off” position and

remove the catheter, or attach a second syringe and repeat aspiration.

c. Blood and thrombus aspirated into the syringe may be filtered for subsequent

laboratory analysis.

To use the basket, wet the mesh to allow fluid flow. Place filter basket in MicroStop fluid

collection basin. Carefully dispense extracted blood into filter basket (blood will go through

filter basket and collect in MicroStop) and filter the extracted blood through the filter basket.

Any fresh, soft emboli and/or thrombi (larger than 70 micron) that have been aspirated

should remain in the filter basket. If necessary, use the second filter basket included in the kit

to continue filtering blood.

13. Remove the ASAP Catheter: if necessary, loosen the rotating hemostasis valve to allow

easy withdrawal of the catheter.

hydrophilic coating 20cm

FREE SLIDING USE

LOCKS TO HOLD VACUUM

English

Advertising