9 ºc – Sper Scientific 850021 Anemometer RH Pen User Manual
Page 6

Unit Selection
1. Turn on the meter by pressing the Power button.
2. Press the Unit button for at least 2 seconds and both the
upper and lower values will start to flash (see below).
Use the ▲ button to switch the upper display unit, and
use the
button to switch the lower display unit. Then
press the
button to accept the selected units.
*
%RH
60.9
25.9 ºC
Value Flashing
Value Flashing
Measurement units:
Anemometer
m/S, Km/h, FPM, mph, Knot
Temperature
ºC, ºF
Humidity
%RH
Dew Point
ºC, ºF
The most recent settings are stored when the power is
turned off, and displayed again when power is turned back
on.
MEASURING PROCEDURE
6
*
Electrical Specifications
Anemometer (Air velocity)
Measure-
ment
Range
Resolution
Accuracy
m/S
0.4 to 30.0 m/
s
0.1 m/S
≤ 20m/S :
± 3% F.S.
> 20 m/S :
± 4% F.S
(all measurements)
Km/h
1.4 to 108.0
km/h
0.1 Km/h
FPM
80 to 5910 ft/
min
1 Ft/min
mph
0.9 to 67.0
mph
0.1 mph
Knots
0.8 to 58.3
knots
0.1 knots
SPECIFICATIONS
11
Temperature
Measuring Range
0 ºC to 50 ºC / 32 ºF to 122 ºF
Resolution
0.1 ºC / 0.1 ºF
Accuracy
±0.8 ºC / 1.5 ºF