Rh calibration – Sper Scientific 800046 Indoor Air Quality Meter User Manual

Page 15

Advertising
background image

- 15 -

RH CALIBRATION

75% Calibration

1. After performing the 33% calibration, plug the sensor

probe into a 75% salt bottle.

2. Press MN/MX/AV o enter the 75% calibration mode.

3. “CAL” and the calibrating value (75.2% if at 25

°

C) blink

on the LCD.


4. The calibration process will be completed in approxi-

mately 60 minutes when “CAL” and humidity stop
blinking.


Note…
Users can also calibrate individually either to 33% or 75%.

To calibrate 33% only, press CAL and exit when 33% cal-
ibration is competed.

To calibrate 75% only, press MODE or DP/WBT within the
first 5 minutes of initializing the 33% calibration process.