Sper Scientific 840021 Light Meter Foot Candle User Manual

Page 4

Advertising
background image

- 4 -

3. MEASUREMENT PROCEDURES

Insert the SENSOR (12) plug into the SENSOR INPUT (8) socket.

Turn the meter on using the ON (2) button.

Remove the lens cap from the PHOTO SENSOR (11).

Press the 200 RANGE (5) button for measurements of 0 ~ 199.9 FC, or
select the 2000 RANGE (6) button for measurements of 200 ~ 1999.9 FC.

If you are uncertain about which range to use, select the highest range.

Range Overload is indicated when "1" appears on the left side of the
display. If this occurs, switch to the highest range or discontinue use.

Point the PHOTO SENSOR (11) toward the source and read the results
on the display. Divide the meter reading by the factor of .0929 to convert
FC results to LUX.

Push the HOLD (4) button to freeze the reading in the display. The word
"HOLD" appears on the LCD. Press the HOLD (4) button again to exit

this function.

Press the OFF (3) button to turn the unit off.

Notes:

Keep the lens cap on the PHOTO SENSOR (11) at all times except when
actually taking a reading.

Fluctuations in the reading are generally due to shadows or fluctuations in
the line voltage. Ambient temperature, drafts and ventilation also affect
the luminous flux output.

Avoid range overload and do not store in areas of high temperature
and / or humidity. Remove the battery for long-term storage.