Calibration line – Sper Scientific 300005 Refractometer - Clinical User Manual

Page 3

Advertising
background image

- 3 -

Operating Procedures
1. With the COVER PLATE open, clean the PRISM

with a soft cloth to avoid scratching the surfaces.

2. Aim the refractometer toward a light source and

rotate the EYEPIECE to obtain the clearest focus.

3. Adjustment of the null (reference point):

A. Open the COVER PLATE.
B. Apply a few drops of pure distilled water onto

the PRISM platform.

C. Close the COVER PLATE.
D. Turn the CALIBRATION SCREW until the

dark and light boundary line coincides with the
bottom line of the scale (calibration line).

4. Carefully dry the prism platform and the cover.
5. Place a few drops of the test solution on the prism

and close the COVER PLATE so the solution
spreads evenly on the prism.

6. Aim the front of the refractometer towards the light

source and focus the eyepiece on the boundary
line of the light and dark hemispheres.

7. The boundary line indicates the concentration of

Urine Specific Gravity and Refractive Index of the
test sample.











8. After use, clean prisms with a cloth and remove

any residue.

9. The temperature of the null reference liquid should

be at the same temperature as the sample solu-
tion. For variations in temperature the null point
should be adjusted once every 30 minutes.

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

13600
80
60
40
20
13500
80
60
40
20
13400
80
60
40
13330

1.050

1.040

1.030

1.020

1.010

1.000

——

——

——

12
11
10
9
8
7
6
5
4
3
2
1
0

W

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

—–

___

ND SERUM URINE Sp.G

JSCP

Calibration

line