3-E. Scale Selection:

Press the SCALE button to toggle

though the scales. The ▼ symbol indicates the active scale.

3-F. High/Low Range Limit Alert

: A red light will flash

when the measurement is outside of the range limit.



Setting the High Range Limit Alert



With the unit on, press the FUNC button as needed until
H.L  Y and the flashing H-Limit icon is displayed.



Press the READ button and the last digit flashes.



Use the ▼ and the ◄ buttons to set the high range limit.



Press the READ button activate the function.



Press the FUNC button as needed to return to the
measurement screen.



Deactivating the High Range Limit Alert



Press the FUNC button as needed until H.L n appears.



Press the Read button to deactivate the function.



Press the FUNC button as needed to return to the
measurement screen.



Setting the Low Range Limit Alert



With the unit on, press the FUNC button as needed until
L.L   Y and the flashing L-Limit icon is displayed.



Press the READ button and the last digit flashes.



Use the ▼ and the ◄ buttons to set the low range limit.



Press the READ button activate the function.



Press the FUNC button as needed to return to the
measurement screen.



Deactivating the Low Range Limit Alert



Press the FUNC button as needed until L.L n appears.



Press the Read button to deactivate the function.



Press the FUNC button as needed to return to the
measurement screen.


3-G. Zero Checking:

Perform Zero Checking if the unit has

sustained a shock, is subjected to temperature changes, or
seems inaccurate.



Fill the prism window with a standard solution. This should
be the same temperature as the solution to be tested.



With the unit on, press the ZERO button. A refractive index
between 1.3329~1.3331 indicates that the unit is accurate.



If the unit is out of range, a flashing crosshair symbol
will be displayed, indicating that recalibration is required.



Press any button to exit this function.