Sper Scientific 300056 Digital Refractometer Custom 1 Scale User Manual

Page 6

Advertising
background image

- 6 -

4. MEASUREMENT

•

Clean the SAMPLE PLATE (1) and
PRISM (3) before and after every use.

•

When the temperature difference between
the ambient temperature and the refracto-
meter is greater than 10°F (5°C), place
the refractometer in the lab and allow it to
adjust to the ambient temperature. Allow
apx. 30 minutes for every 5°F (10°C)
difference in temperature.

•

Press the READ (9) button to turn the unit
on and display the temperature. This unit
features Automatic Temperature
Compensation (ATC) within the range of
41~104°F (5~40°C), eliminating the need
for manual temperature correction.

•

Perform a calibration (pg 6) before the
day's first use or after extended daily use.

•

Fill the PRISM (3) with about 1.0ml of the
sample solution.

•

Close the PRISM COVER (2).

•

Press the READ (9) button immediately to
obtain a reading. Delays may cause the
sample to evaporate.

•

Within 1 second, the value is displayed.

•

When the reading is under-range, the low
indicator (LLL) is displayed. The high
indicator (HHH) is displayed when the
reading is over-range.

- 7 -

Models with Multiple Scales

•

Follow the measurement instructions
(page 6), then press the SCALES (8)
button and hold the button for 1 second to
cycle through the options.

•

"- - -" is displayed and the ACTIVE
SCALE
(7) indicator "▼" points to the
selected scale.

•

With the desired scale selected, press the
READ (9) button to obtain the results.

Notes and Precautions

•

The refractometer will automatically shut
off after about 3 minutes of inactivity.

•

After each use, clean the SAMPLE
PLATE
(1) and PRISM (3) with distilled
water on a soft cloth, and wipe dry.

•

If a film develops on the PRISM (3) it will
repel test samples and affect readings.
If this occurs, clean with a mild detergent.

•

The prism is easily scratched, never use
abrasive cleaning materials.

•

Never submerge.

•

Avoid spills around the unit.

•

Avoid prolonged use or storage in
extreme temperatures.

•

Do not use in corrosive environments.