Sper Scientific 840015 Sound Meter Type 1 User Manual

Page 6

Advertising
background image

6

Frequency A Weighting C Weighting

Tolerance

(IEC 61672

Class 1)

31.5 Hz

-39.4 dB

-3 dB

± 2.0 dB

63 Hz

-26.2 dB

-0.8 dB

± 1.5 dB

125 Hz

-16.1 dB

-0.2 dB

± 1.5 dB

250 Hz

-8.6 dB

0 dB

± 1.4 dB

500 Hz

-3.2 dB

0 dB

± 1.4 dB

1 kHz

0 dB

0 dB

± 1.1 dB

2 kHz

+1.2 dB

-0.2 dB

± 1.6 dB

4 kHz

+1 dB

-0.8 dB

± 1.6 dB

8 kHz

-1.1 dB

-3 dB

+2.1 dB

-3.1 dB

6. FREQUENCY WEIGHTING (A and C) CHARACTERISTICS

Tolerance

(ANSI S1.43

Type 1)

± 1.5 dB

± 1.0 dB

± 1.0 dB

± 1.0 dB

± 1.0 dB

± 1.0 dB

± 1.0 dB

± 1.0 dB

+1.5 dB

-6.0 dB

12.5 kHz

-4.3 dB

-6.2 dB

+3.0 dB

-6.0 dB

+3.0 dB

-6.0 dB

16 kHz

-6.6 dB

-8.5 dB

+3.5 dB

-17.0 dB

+3.0 dB
-no limit

Time Weighting

Max. response ref.

continuous signal

Tolerance

(IEC 61672 Type 1)

F (Fast)

-1.0 dB

+ 1.0 db

S (Slow)

-4.1 dB

± 1.0 db

Tested with signal on 1000 Hz/94 dB.