Sper Scientific 800005 Thermometer Advanced w/RS232 User Manual
Page 4

3
Accuracy: at ( 23 ± 5°C )
Range Accuracy
K-Type -200°C ~ 1370°C
±(0.1% reading + 0.7°C)
J-Type -200°C ~ 760°C
±(0.1% reading + 0.7°C)
K-Type -328°F ~ 2498°F
±(0.1% reading + 1.4°F)
J-Type -328°F ~ 1400°F
±(0.1% reading + 1.4°F)
Temperature Coefficient:
For ambient temperatures from 0°C ~ 18°C and
28°C ~ 50°C, for each °C ambient below 18°C or
above 28°C add the following tolerance into the
accuracy spec.
0.01% of reading + 0.03°C
( 0.01% of reading + 0.06°F )
Note:
The basic accuracy Specification does not include the
error of the probe please refer to the probe accuracy
specification for additional details.
Sample Rate: 3.3 times per second
Dimension: 184
Ч
64
Ч
30mm
Weight: 210g Approx.
Accessory:
K Type Bead Probe, Battery, Carrying Case,
Instruction Menu.
Option:
Soft Ware Package(Program, RS
232
Connection
Cable), AC Adapter
Power requirement:
9 Volt Battery, NEDA 1604 or JIS 006P or IEC6F22