Table 206

Recommended separation distances between portable and mobile RF

communications equipment and the Posey Bed 8070

The Posey Bed 8070 is intended for use in an electromagnetic environment in which radiated RF

disturbances are controlled. The customer or the user of the Posey Bed 8070 can help prevent

electromagnetic interference by maintaining a minimum distance between portable and mobile RF

communications equipment (transmitters) and the Posey Bed 8070 as recommended below, according to

the maximum output power of the communications equipment.

For transmitters rated at a maximum output power not listed above, the recommended separation

distance d in meters (m) can be estimated using the equation applicable to the frequency of the

transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the

transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

Separation distance according to frequency of transmitter

m

Rated maximum

output power

of transmitter

W

0.01

0.1

1

10

100

150 kHz to 80MHz

d = [3.5/V1]√P

0.12

0.37

1.17

3.69

11.67

80 MHz to 800 MHz

d = [3.5/E1]√P

0.12

0.37

1.17

3.69

11.67

800 MHz to 2.5 GHz

d = [7/E1]√P

0.23

0.74

2.33

7.39

23.33