User's Guide ________________________________________________________
14 _____________________________________________________M210185EN-C
0606-166
Figure 3
Pour a correct dose of ion exchanged
water into the chamber; dry the
measurement cup
3.
Sprinkle the contents of a salt package (or measure the
salt according to the table) in small quantities into the
chamber, stirring constantly. When measuring with the
measurement cup, make sure that the cup is clean and
dry. Rinse and dry the cup after every use.
LiCl
15 g or 18 ml
MgCl2
30 g or 30 ml
NaCl
20 g or 15 ml
K2SO4
30 g or 20 ml