Two point calibration, Head – Vaisala HMM213 User Manual

Page 38

Advertising
background image

USER'S GUIDE____________________________________________________________________

>CT -70 0 101 -0.5
>

Use an accurate and calibrated reference only.

CT Temperature Calibration of the
Additional Sensor Head

Two Point Calibration

CTA <cr>

Using this command the transmitters can be calibrated against an
accurate reference, such as a Pt 100 simulator. A two-point calibration is
performed as follows:

>CTA <cr>
T : 23.19 Ref1 ?

23.3 <cr>

Press any key when ready ...
T : 101.42

Ref2 ?

101 <cr>

If the stabilization of the sensor to the temperature of the reference needs
to be monitored, the measurement output can be repeated with <cr> at
Ref1 and Ref2:

>CTA <cr>
T : 23.19 Ref1 ?

<cr>

T : 23.20 Ref1 ?

<cr>

.
.
.
T : 23.22 Ref1 ?

23.3 <cr>

Press any key when ready ...
T : 101.42

Ref2 ?

101 <cr>

38 ____________________________________________________________________ U339EN-1.3