One point calibration, Gain calibration – Vaisala HMM213 User Manual

Page 39

Advertising
background image

Chapter 6 _______________________________________________________________ Calibration

One Point Calibration

CTA xx.xx yy.yy

xx.xx =

current temperature

yy.yy =

offset correction

Example of performing the offset calibration in one reference
temperature:

>CTA 21.2 2.3
>

Use an accurate and calibrated reference only.

Gain Calibration

CTA xx 0 yy.yy zz.zz

xx =

temperature in which there is no changes in offset

yy.yy =

current temperature

zz.zz =

correction (T

ref

-T

measured

)

Example of performing the gain calibration in one reference point
without changing the offset at 0 °C:

>CTA 0 0 101 -0.5
>

Example of performing the gain calibration in one reference point
without changing the offset at -70 °C:

>CTA -70 0 101 -0.5
>

Use an accurate and calibrated reference only.

VAISALA _______________________________________________________________________ 39