SCALING THE ANALOG OUTPUT VIA THE SERIAL BUS

Serial communication settings

The analog output scaling can be done using the serial bus and computer with suitable terminal software.
Connect the DMT242 to a serial bus via the RS-232 interface by using the following settings: Baud rate:
2400, Parity: none, Data bits: 8, Stop bits: 1
. The serial cable (DMT242RS) can be ordered from Vaisala.

Scaling of the dewpoint output

Scale the dewpoint parameter by giving the command ascl xx yy<ENTER> where xx = the low limit
(

C or ºF) and yy = the high limit (C or ºF). The factory setting for serial measurement unit is

Centigrade (

C). To change the unit for Fahrenheit (ºF) use command unit xx <ENTER> where xx = n

for non-metric (ºF) and xx = m for metric (

C) units.

ACCESSORIES

Order Code

Description

HM47280 Stainless

steel sintered filter

DMT242RS

RS-232 serial line cable for PC (with female D connector)

210662

1/2" NPT adapter