User's Guide _______________________________________________________________________

150 __________________________________________________________________ M210704EN-G

Error States

In error state the affected quantities are not measured and the output is
shown as follows:

- Analog channel outputs 0 mA or 0 V (you can use the serial line

command AERR or display/keypad to change this fault indication
value, see section Analog Output Fault Indication Setting on page
127).

- Serial port outputs stars "***" instead of measurement data.

- Display shows "----" instead of measurement data.

- Cover LED is blinking.

- Error indicator is shown on the display.

0706-017, 1103-069

Figure 80

Error Indicator and Error Message

Number refers to Figure 80 above:

1

=

Error indicator

The error indicator disappears when the error state is over and you have
checked the error message. Press the INFO button to display the error
message.

You can also check the error message via the serial interface by using the
command ERRS. In case of constant error, please contact Vaisala
Technical Support. See section Technical Support on page 152.