Operation, Transmitter start-up, Chapter 5 – Vaisala DPT145 User Manual

Page 30: Chapter 5, operation

Advertising
background image

User's Guide _______________________________________________________________________

30 ___________________________________________________________________ M211371EN-D

CHAPTER 5

OPERATION

This chapter contains information that is needed to operate the DPT145.

Transmitter Start-Up

When the transmitter is connected to power supply, there is a delay of

five seconds as the transmitter starts up. After the start-up is complete,

the transmitter will behave according to the configured serial mode:

- In STOP mode, the transmitter outputs the transmitter model and

software version.

- In RUN mode a measurement output starts immediately.
- In POLL mode the transmitter does not output anything after power-

up.

- In MODBUS mode, the transmitter does not output anything after

power-up.

For a description of the modes, see section Set Serial Line Operating

Mode on page 41.

For a description of how the diagnostic functions of the sensor affect the

availability of live measurement at start-up, see section DPT145 Start-up

Sequence on page 20.

Advertising