Vaisala PTB200 User Manual

Page 19

Advertising
background image

PTB 200

18 (28)

Digital Barometers

SSD/Operating Manual

24th February 1993

PTB200-O0284-1.1

SLEEP

Setting and resetting the sleep mode

SLEEP ON<cr>
SLEEP OFF<cr>

The SLEEP command is used to set and reset the software controlled
sleep mode which cuts down the power consumption by over 60 %. In
the sleep mode the current consumption is below 10 mA when no out-
put is required and about 25 mA during output. By activating the sleep
mode the barometer automatically falls asleep when no output is
required. Even then the barometer is ready to accept commands and
execute them. However, because the barometer needs to wake up, the
time from a command to output is a bit longer in SLEEP mode than in
normal mode.

?

List of basic software settings

?<cr>

Entering ? outputs a list of basic software settings, e.g.

Baud Parity Data Stop 1200 E 7 1
Output format

\PPPP.PP\ \uuuu\\r\n

Pressure unit

hPa

Temperature unit

’C

Sending mode

STOP

Address

0

Sleep mode

OFF

Measurement time

64 x 25 milliseconds

Filter

OFF / FAST

Output interval

0 s

Calibration day code

92261

Offset drift comp.

ON

Offset drift correction

-0.056 hPa

Multipoint correction

ON
Reading

Correction

499.64

0.024

599.79

0.013

699.92

0.017

800.04

0.021

900.18

0 017

1000.33

0.023

1100.43

0.027

Advertising