Technical data, Technical specifications, Operating range – Vaisala PTB210 (analog) User Manual
Page 21: Accuracy, Technical specifications operating range, Chapter 5
Advertising

Chapter 5 _____________________________________________________________ Technical Data
VAISALA________________________________________________________________________ 19
CHAPTER 5
TECHNICAL DATA
Technical Specifications
Operating Range
Pressure range (order specified)
500 ... 1100 hPa
800 ... 1060 hPa
600 ... 1060 hPa
900 ... 1100 hPa
Operating temperature range
-40 ... +60
°C
Humidity range
non condensing
Accuracy
Non linearity*
± 0.20 hPa
Hysteresis*
± 0.05 hPa
Repeatability*
± 0.05 hPa
Calibration uncertainty**
± 0.15 hPa
Total accuracy (20
°C)***
± 0.30 hPa
Temperature dependence
(-40 ... +60
°C)****
± 0.50 hPa
Total accuracy***
(-40 ... +60
°C) ± 0.60 hPa
Advertising