Vaisala WAT12 User Manual

Page 42

Advertising
background image

User's Guide ________________________________________________________

40 _____________________________________________________M210309EN-A

Property

Description / Value

Signal cable

4 wires minimum
(VIN+, VIN-, DOP, SOP)

Max. loop resistance
(incl. cable and
receiver's input
resistance)

1800 ohm for 5 mA loop
900 ohm for 10 mA loop
450 ohm for 20 mA loop

Full scale options:
for direction
for speed

0 ... 360° / 0 ... 540°
0 ... 51.2 m/s / 0 ... 76.8 m/s

Resolution:
for direction
for speed

±2.8

°

0.1 m/s

Updating interval:
for direction
for speed

8 times a second (with no averaging)
Once a second (averaged over
previous 1 s period)

Output accuracy

Better than 1 % of full scale

Accuracy with loop
resistance:
RL= 100 Ω

RL= 20 ... 2000 Ω

±0.2 % of full scale at +25 °C
±0.8 % of full scale at full temp. range

Pulse-mode power
feed:
to the vane

to the anemometer

Nominally 490 µs wide pulse,
applied every 125 ms
Nominally 70 µs wide pulse,
applied every 0.49 ms .

Testing references:
for direction
for speed

000° / 008° / 357°
0.0 m/s / 0.8 m/s / 51.2 m/s

Dimensions:
printed circuit board
junction box

114 × 69 mm

2

125 (w) × 80 (h) × 57 (d) mm

Cross arm length

800 mm

Ingress protection

IP65

Mounting

To a Ø 60 mm pole mast

Weight

1.5 kg

Material:
cross arm
junction box

AI anodized
AI painted gray

Temperature range:
operating
storage

-55 ... +55 °C
-60 ... +70 °C

Humidity

0 ... 100 %RH

Advertising