Ashcroft, Previous type 1009 duralife, Calibration procedure – pressure range – Ashcroft 1036 Gauge with 1037 Fitting User Manual

Page 13

Advertising
background image

Step 1. With the dial off, install pointer at 9 o’clock “lightly,”

Figure 3.

Step 2. Go to full scale pressure…rotate span block with

tool until pointer rests at 6 o’clock.

Step 3. Go to zero pressure (9 o’clock)…if pointer has not

moved away from start point, go to Step 4. If pointer
has moved, repeat Step 1 until span is correct.

Step 4. Install dial with screws snug.

Dial Screws

Zero Box

Full Scale Mark

Mid Scale Mark

Figure 4 –

ASHCROFT

®

Previous Type 1009 Duralife

®

Calibration Procedure – Pressure Range

© 2010 Ashcroft Inc., 250 East Main Street, Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0499, www.ashcroft.com
All sales subject to standard terms and conditions of sale. I&M008-10077 9/01 AMR 05/10

Step 5. Install pointer centered in zero box, Figure 4.

Step 6. Go to full scale pressure…check that pointer is

within 1% of full scale mark. If not, remove pointer
and dial and return to step 1, Figure 4.

Step 7. Go to mid-scale pressure…rotate dial until mid-

scale mark is aligned with pointer, Figure 4.

Step 8. Tighten dials screws and stake on pointer.

Step 9. Check zero and full scale. Reassemble window,

gasket and ring.

Notes: See page 10 for any tools required to calibrate.

For model s pr oduced pr i or t o
Sept ember 2008 f or 2

1

2

˝ ver si on and

December 2008 f or 3

1

2

˝ ver si on.

Back of gauge will have a date code sticker.

Span Block

Decrease

Increase

Start Point

Span

Figure 3 –

13