THIS TEST GAUGE IS PROVIDED WITH A MICROSPAN™ ADJUSTMENT

TO SIMPLIFY CALIBRATION. THE FLOW CHART BELOW OUTLINES THE

RECOMMENDED CALIBRATION PROCEDURE

Installation and Maintenance Instructions for

ASHCROFT

®

Type 1082 Test Gauge

Calibration Procedure – Pressure Range

CENTER DIAL
OVER PINION

CHECK

ACCURACY

AT 0-25-50-75-100%

MICRO

SPAN

COARSE

SPAN

GO TO

0%

MICROMETER

SPAN

COARSE

SPAN

>0.2 <1.5%

>1.5%

<0.2%

>0.2<1.5%

SET POINTER

AT 0%

GO TO

100%

CHECK

ACCURACY

>0.2<1.5%

TEMPERATURE
COMPENSATOR

< – LESS THAN
> – GREATER THAN

MICROSPAN
ADJUSTMENT

–TIGHTEN SCREW TO

INCREASE SPAN
1 TURN – APPROX.

1

2

%

COARSE SPAN
ADJUSTMENT

–LOOSEN TWO SCREWS

–MOVE SLIDE INWARD

TO INCREASE SPAN
(.004 IN. – APPROX. 1%)

–TIGHTEN TWO SCREWS

>1.5%

>1.5%

<0.2%

<0.2%

GO TO

100%

CHECK

ACCURACY

GO TO

0%

ROTATE

MOVEMENT

SET POINTER

AT 0%

GO TO

50%

CHECK

ACCURACY

MICRO

SPAN

>0.2%

COARSE

LINEARITY

NOTE

<0.2%

SET POINTER

AT 100%

GO TO

0%

CHECK

ACCURACY

COARSE

SPAN

<0.2%

>1.5%

SET POINTER

AT 100%

GO TO

0%

CHECK

ACCURACY

ROTATE MOVEMENT
COUNTER-CLOCKWISE
TO REDUCE READING
AT MIDSCALE