Campbell Scientific 253-L and 257-L (Watermark 200) Soil Matric Potential Sensors User Manual

Page 27

Advertising
background image

253-L and 257-L Soil Matric Potential Sensors

;{CR10X}

*Table 1 Program
01: 1.0000 Execution

Interval

(seconds)


;Measure soil temperature with 107 sensor
1: Temp (107) (P11)
1: 1

Reps

2: 1

SE

Channel

3: 1

Excite

all

reps

w/E1

4: 1

Loc [ Tsoil_C ]

5: 1.0

Multiplier

6: 0.0

Offset


;Measure 257 block resistance
2: AC Half Bridge (P5)
1: 1

Reps

2: 14

250 mV Fast Range

3: 2

SE

Channel

4: 2

Excite

all

reps

w/Exchan

2

5: 250

mV

Excitation

6: 2

Loc [ kOhms ]

7: 1

Multiplier

8: 0

Offset


;Convert Half Bridge reading to kOhms
3: BR Transform Rf[X/(1-X)] (P59)
1: 1

Reps

2: 2

Loc [ kOhms ]

3: 1

Multiplier

(Rf)


;Calculate dT = T -21
4: Z=X+F (P34)
1: 1

X Loc [ Tsoil_C ]

2: -21

F

3: 4

Z Loc [ CorFactr ]


;Calculate (0.018 * dT)
5: Z=X*F (P37)
1: 4

X Loc [ CorFactr ]

2: 0.018

F

3: 4

Z Loc [ CorFactr ]


;Calculate (1 - (0.018 * dT))
6: Z=X+F (P34)
1: 4

X Loc [ CorFactr ]

2: -1

F

3: 4

Z Loc [ CorFactr ]


7: Z=X*F (P37)
1: 4

X Loc [ CorFactr ]

2: -1

F

3: 4

Z Loc [ CorFactr ]

21