Campbell Scientific 253-L and 257-L (Watermark 200) Soil Matric Potential Sensors User Manual

Page 31

Advertising
background image

253-L and 257-L Soil Matric Potential Sensors

;{CR10X}
01: 30.0000 Execution

Interval

(seconds)


;Turn on AM16/32
1: Do (P86)
1: 41

Set

Port

1

High


;Loop to measure five 107 probes and five 253's
2: Beginning of Loop (P87)
1: 0

Delay

2: 5

Loop

Count


;Advance to next multiplexer channel

3: Do (P86)

1: 72 Pulse Port 2


;10 msec delay to allow switch to settle

4: Excitation with Delay (P22)

1: 1

Ex

Channel

2: 0

Delay W/Ex (0.01 sec units)

3: 1

Delay After Ex (0.01 sec units)

4: 0

mV

Excitation


;Measure soil temperature

5: Temp (107) (P11)

1: 1

Reps

2: 1

SE

Channel

3: 1

Excite all reps w/E1, 60Hz, 10ms delay

4: 1

-- Loc [ T107_C_1 ]

5: 1.0

Multiplier

6: 0.0

Offset


;Measure 253 sensor

6: AC Half Bridge (P5)

1: 1

Reps

2: 14

250 mV Fast Range

3: 2

SE

Channel

4: 2

Excite

all

reps

w/Exchan

2

5: 250

mV

Excitation

6: 11

-- Loc [ kOhms_1 ]

7: 1

Multiplier

8: 0

Offset


;Convert Half Bridge reading to resistance (k-Ohm)

7: BR Transform Rf[X/(1-X)] (P59)

1: 1

Reps

2: 11

-- Loc [ kOhms_1 ]

3: 1

Multiplier

(Rf)


;Apply nonlinear equation from 5.3.2

8: Z=X*Y (P36)

1: 1

-- X Loc [ T107_C_1 ]

2: 1

-- Y Loc [ T107_C_1 ]

3: 6

-- Z Loc [ WP_kPa_1 ]

25

Advertising