Campbell Scientific AM16/32A Multiplexer User Manual

Page 35

Advertising
background image

AM16/32A Relay Analog Multiplexer

COM EVEN H

COM EVEN L

EVEN H

EVEN L

SENSOR SHIELDS

COM

"4 X 16" Mode

COM ODD H

COM ODD L

CU

CO

CU

CO

ODD H

ODD L

CU

CO

CU

CO

/

0


21X/

CR7



CR10(X)

CR3000/

CR800/CR850/

00/

CR23X/CR10

CR5000

H H

H

L L

L

H H

H

L L

L

G

FIGURE 17. Differential Thermocouple Measurement with Reference Junction at the

Datalogger.

COM EVEN H

COM EVEN L

EVEN H

EVEN L

SENSOR SHIELDS

COM

"4 X 16" Mode

COM ODD H

COM ODD L

107

ODD H

ODD L

CU

CO

CU

CO

CU

CU

CU

CU

/

0

CR23X/

CR800/
CR850/

CR1000


21X/

CR7



CR10(X)


CR3000/

CR5000

H H H H

L L L L

EX E E VX

H H H H

AG

H H H H

L L L L

G

FIGURE 18. Differential Thermocouple Measurement with Reference Junction at the

AM16/32A.

If a mix of TCs and other sensor types are multiplexed through the
AM16/32A, it is generally best to locate the reference junction on the
AM16/32A, as shown in Figure 18.

AM16/32A Reference - An external reference, usually a thermistor,
can be located at the AM16/32A, as shown in Figure 18. This
approach requires an additional single-ended datalogger input to
measure the reference. Position the reference next to the COM
terminals and, when practical, measure the thermocouples on SETs
that are in close proximity to the COM terminals in order to minimize
thermal gradients.