3 power conservation, 2 measuring multiple sdi-12 sensors, Troubleshooting and maintenance – Campbell Scientific CS215 Temperature and Relative Humidity Probe User Manual

Page 22: 1 troubleshooting, Power conservation, Measuring multiple sdi-12 sensors, Troubleshooting

Advertising
background image

CS215 Temperature and Relative Humidity Probe

8.1.3 Power Conservation

The CS215 draws less than 70 µA of current between measurements. In most

applications, this is insignificant compared to the datalogger and other power

uses, so the sensor can be permanently connected.

In low-power applications, battery power can be conserved by turning the 12 V

supply to the CS215 on just before the measurement (allowing a ‘warm-up’

time of at least 100 ms) and then turning it off afterwards. If available, the

switched 12 V output of the datalogger can be used.

8.2 Measuring Multiple SDI-12 Sensors

Up to ten CS215s or other SDI-12 sensors can be connected to a single

datalogger control port. Each SDI-12 device must have a unique SDI-12

address between 0 and 9.

9. Troubleshooting and Maintenance

All factory repairs and recalibrations require a returned material

authorization (RMA) and completion of the “Declaration of

Hazardous Material and Decontamination” form. Refer to the

Assistance

page at the beginning of this manual for more

information.

9.1 Troubleshooting

Symptom: Temperature is reported as

–9999 or NAN, and relative humidity is

reported as

0 or as an unchanging value.

Recheck wiring. Verify the green wire is connected to the control port

specified by the

SDI12Recorder() instruction. Verify the red wire is

connected to a 12V terminal.

Check the voltage to the sensor with a digital voltage meter. If a switched 12V

terminal is used, temporarily connect the red wire to a 12V terminal (non-

switched) for test purposes.

Verify the probe’s SDI12 address matches the address entered for the

SDI12Recorder() instruction. The default address is 0. The address can be

verified or changed with the commands described in Appendix D, SDI-12

Sensor Support.

Remove the filter tip and verify that the sensing element has been installed

with the proper orientation as described in Section 9.4, Sensor Element

Replacement.

Symptom: Incorrect temperature or relative humidity is reported.

If using the SW12 terminal to power the sensor, verify the program is allowing

a warm-up time of at least 100 ms.

Check to see if the filter tip has been contaminated. Replace the filter tip (pn

18142), or clean with distilled water as needed.

NOTE

12

Advertising