Campbell Scientific EC150 CO2 and H2O Open-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 18

Advertising
background image

EC150 CO

2

/H

2

O Open-Path Gas Analyzer

Gas analyzer

Measurement precision

iii

CO

2

density:

0.2 mg·m

–3

(0.15 µmol·mol

–1

)

H

2

O density:

0.004 g·m

–3

(0.006 mmol·mol

–1

)

Factory calibrated range

CO

2

:

0 to 1000 µmol·mol

–1

H

2

O:

0 to 72 mmol/mol (37°C dewpoint)

Temperature:

30° to 50°C

Barometric pressure:

70 to 106 kPa

CO

2

performance

Zero max drift

iv

:

±0.55 mg·m

–3

·°C

–1

(± 0.3

μmol·mol·°C

–1

)

Gain drift:

±0.1% of reading·°C

–1

(maximum)

Sensitivity to H

2

O:

±1.1 x 10

–4

µmol CO

2

·mol

–1

H

2

O (max)

H

2

O performance

Zero max drift

iv

:

±0.04 g·m

–3

·°C

–1

(± 0.05 mmol·mol

–1

·°C

–1

)

Gain drift:

±0.3% of reading·°C

–1

(maximum)

Sensitivity to CO

2

:

±0.1 mol H

2

O·mol

–1

CO

2

(maximum)

CSAT3A sonic measurement precision

v

u

x

:

1.0 mm·s

–1

u

y

:

1.0 mm·s

–1

u

z

:

0.5 mm·s

–1

Sonic temperature:

0.025°C

CSAT3A sonic accuracy

vi

Offset error

u

x

, u

y

:

< ±8 cm·s

–1

u

z

:

< ±4 cm·s

–1

Gain error

Wind vector ±5° horizontal: < ±2% of reading

Wind vector ±10° horizontal: < ±3% of reading

Wind vector ±20° horizontal: < ±6% of reading

iii

noise rms, assumes:

o 25°C

o 85 kPa

o 14 g·m

-3

H

2

O

o 597 mg·m

-3

CO

2

o 25 Hz bandwidth

iv

30° to 50°C

v

noise rms

vi

assumes:

o −30° to +50°C

o wind speed <30 m·s

-1

o azimuth angles between ±170°

8