EC150 CO
2
/H
2
O Open-Path Gas Analyzer
bracket (it should slide into place and be able to securely hang from
the bracket), and retightening the bolts (see FIGURE 6-8).
FIGURE 6-8. Exploded view of mounting the EC100 enclosure
5. Remove the EC100 enclosure desiccant from the plastic bag and put it
back in the mesh pocket of the enclosure.
6. Adhere the humidity indicator card to the inside of the door of the
enclosure.
6.2.5 Install the EC150 Temperature Probe
The temperature probe should be mounted such that it measures at the same
height as the sample volume of the EC150 and the CSAT3A.
1. Attach the R.M. Young 41303-5A 6-plate solar radiation shield (pn
4020) to the mast with the included u-bolt.
2. Insert the end of the temperature probe into the hole on the bottom of
the shield, see FIGURE 6-10.
3. Tighten screws to hold the probe in place.
18