EC150 CO

2

/H

2

O Open-Path Gas Analyzer

Avoid crashing the arms of the sensors together. The arms

of the analyzer should slide in between the claws of the

CSAT3A; the sonic head may need to be loosened and

repositioned to do this.

10. Tighten bolts and check that the analyzer is oriented vertically such

that the label is right-side-up and the upper arm (source) is directly

above the lower arm (detector).

11. If the assembly is not level, slightly loosen the bolt that holds the

mounting bracket on the leveling mount and adjust the assembly until

the leveling bubble on the top of the CSAT3A head is within the

bullseye. Retighten the bolt.

FIGURE 6-5. Exploded view of mounting CSAT3A and EC150

CAUTION