Specifications – Campbell Scientific SDM-INT8 8-Channel Interval Timer User Manual

Page 9

Advertising
background image

SDM-INT8 8 Channel Interval Timer

2. Specifications

Operating voltage: 9.6 V to 16 V DC

Current drain: 13 to 20 mA active, 400 microamp quiescent

Environmental: -25 to +50 degrees Celsius 0 to 90% RH (non-condensing)

Number of channels: 8

Maximum timing measurement: 16.7 seconds

Resolution: ± 1 microsecond

Dimensions: 8 x 5 x 1 in (13 x 20 x 2 cm)

Weight: 1.4 lbs (635 g)

Input voltage option per channel: high level, low level

High Level Voltage Input:

- Minimum pulse width: 2 microseconds

- Signals edges:

rising: transition from < 1.5 to > 3.5 volts

falling: transition from >3.5 to<1.5 volts

Maximum input voltage: 20 volts

- Maximum frequency:

5.1 kHz when using Averaging Options

10 kHz when Capturing All Events

The Low Resolution Frequency function allows higher frequencies to be
measured if it is used on all programmed channels with Execution Interval
Averaging. Maximum frequency is dependent on the number of channels
programmed, as shown below:

No. of Channels Max Freq (kHz)
1

42.5

2

17.5

3

11.0

4

8.6

5

5.2

6

4.8

7

4.5

8

4.28

Low Level Voltage Input:

- Minimum AC voltage: 20 millivolts RMS

- Input Hysteresis: 11 millivolts

- Maximum AC voltage: 20 volts RMS

- Minimum frequency: 1 Hz

- Maximum frequency:

3