WindSonic Two-Dimensional Sonic Anemometer

7.2 Wiring

7.2.1 Serial Wiring (COMn)

CRBasic dataloggers (CR800 series, CR1000, and CR3000) support serial
communications with dedicated UART hardware on the datalogger control
ports. Two control ports can be configured as a single communications
(COMn) port. The WindSonic1 serial interface uses four wires as shown in
TABLE 7-2.

TABLE 7-2. WindSonic1 to Datalogger Connections

Description Color

CRBasic

Datalogger

WindSonic RxD

Green COMn Tx

WindSonic TxD

White COMn Rx

Power Red

12

Vdc

Serial/power reference Black G

Shield Clear

G

The maximum cable length that can be used with a RS-232
interface depends on the baud rate, the nominal resistance of the
wire, the capacitance between conductors, and the capacitance
between the conductors and the shield. According to the
Electronic Industries Association RS-232D standard, a rough
rule of thumb is to limit RS-232 cable lengths to 15.24 m (50
feet) or less at 9600 baud.

NOTE

7.2.2 Serial Wiring (SDM-SIO1)

CRBasic dataloggers support serial communications using the SDM-SIO1
peripheral. The SDM-SIO1 converts RS-232 signals into Synchronous Device
for Measurements (SDM). SDM is a Campbell Scientific digital
communications protocol used between Campbell Scientific dataloggers and
SDM peripherals. At a 1 Hz measurement rate, a maximum of 4 WindSonic1s
can be measured by a datalogger. TABLE 7-3 describes the connections
between a WindSonic1 and SDM-SIO1.

TABLE 7-3. WindSonic1 to SDM-SIO1 Connections

Description Color

SDM-SIO1

WindSonic RxD

Green

TX-Z

WindSonic TxD

White

RX-A

Power Red

+12V

Serial/power reference

Black

G

Shield Clear

G