Gpib operation ml24xxa native commands – Anritsu ML2430A User Manual

Page 146

Advertising
background image

Query:

OBDST? <port>

Returned

String:

OBDST <port>,<units>,<val>

OBMD

BNC output mode select

BNC

Syntax:

OBMD <port>,<mode>

port:

1 or 2

mode:

'OFF' (output set to ground) port 1 or 2
'AOUT' (analog scaled output) port 1 or 2
'PASS/FAIL' (pass/fail) port 1 or 2
'SIGA' (signal output sensor A) port 1 only
‘LVLA1’ Signal channel range 1 amplifier output for sensor A
‘LVLA2’ Signal channel range 2 amplifier output for sensor A
‘LVLB1’ Signal channel range 1 amplifier output for sensor B
‘LVLB2’ Signal channel range 2 amplifier output for sensor B
'ACMOD' (AC mod output) port 1 only
'RFB' (RF blanking while zeroing) port 2 only
'SIGB' (signal output sensor B) port 2 only

Remarks:

Changes the type of output selected for the BNC outputs.

Query:

OBMD? <port>

Returned

String:

OBMD <port>,<mode>

OBPL

BNC pass/fail pass level

BNC

Syntax:

OBPL <port>,<level>

port:

1 or 2

level:

HIGH (TTL high is PASS)
LOW (TTL low is PASS)

Remarks:

Selects the PASS level for the Pass/fail type of output.

Query:

OBPL? <port>

Returned

String:

OBPL <port>,<level>

OBVSP

BNC analog output stop voltage scale

BNC

Syntax:

OBVSP <port>,<val>

ML2430A OM

6-59

GPIB OPERATION

ML24XXA NATIVE COMMANDS