Gpib operation hp 437b emulation commands – Anritsu ML2430A User Manual

Page 194

Advertising
background image

FA

Auto average

Syntax:

FA

Remarks:

Automatic Filter on. Allows the system to automatically select the filter
used to reduce the jitter in the display.

Related

Commands:

FM, FH

FH

average hold

Syntax:

FH

Remarks:

Hold filter sets the filter mode to Manual from Auto, but retains the auto fil-
ter setting. This function is the same as the AVGM command.

Related

Commands:

FM, FH, FA

FM

Set average value

Syntax:

FM<val>EN

val:

1 to 512

Remarks:

Sets the filter length for the averaging of sensor data. For HP 437B emula-
tion, the command accepts 1 to 512 in 2-to-the-power steps. For example,
1, 2, 4, 8, 16,...256, 512.

Related

Commands:

FH, FA

FR

Frequency of the input signal

Syntax:

FR<val><units>

val:

units:

GZ (GHz)
MZ (MHz)
KZ (KHz)
HZ (Hz)
EN (Hz)

Remarks:

Sets the frequency of the input signal so that the correct cal factor is used.

ML2430A OM

6-107

GPIB OPERATION

HP 437B EMULATION COMMANDS

Advertising