Ml4803a emulation commands gpib operation – Anritsu ML2430A User Manual

Page 173

Advertising
background image

CCA

Clear the calfactor value to zero.

CDJ

Perform a CAL 0 dBm.

Remarks:

During the cal 0 dBm sequence, the CAL execution bit in the status byte is
set. When the CAL operation is completed, the CAL execution bit is
cleared.

COF

Clear the offset value to zero.

COS

Turn ON the 50 MHz, 0 dBm RF calibrator output.

CRF

Clear the reference value to zero.

CST

Turn OFF the 50 MHz, 0 dBm RF calibrator output.

DBM

Sets the display channel units to dBm.

DBR

Sets the display channel units to dB’s and takes the relative
value.

Remarks:

The relative value is stored as the reference data. The reference value can
be independently changed with the GPIB command REF.

EMUL

GPIB emulation mode

Syntax:

EMUL <mode>

mode:

ML24XX (Anritsu ML2430A Series native mode)
HP436A (Hewlett-Packard)
HP437B (Hewlett-Packard)
HP438A (Hewlett-Packard)
ML4803 (Anritsu ML4803A Series)

Remarks:

Set the GPIB emulation to emulate other types of power meters. This com-
mand is available in any emulation mode, and resets the whole GPIB inter-
face when the emulation mode is changed.

6-86

ML2430A OM

ML4803A EMULATION COMMANDS

GPIB OPERATION