9 percentage mode, 1 storing a 100% mass – A&D Digital Platform Scales HW-10KGV User Manual

Page 25

Advertising
background image

HW-G Instruction Manual Page 23

HW-G OZIM v1a

9 Percentage Mode

The percentage mode is to display a mass value in the unit of “%”.

It is necessary to store a 100% mass value in advance to use this function.

9.1 Storing a 100% Mass

Step 1

Press the

MODE

switch to display the unit %.

Step 2

Press the

SET

switch to enter the mode that

stores a 100% mass.

Step 3

If necessary, with nothing on the pan press
the

ZERO

switch to display zero.

Step 4

Gently place the 100% mass onto the pan.
Press the

ENTER

switch to store the mass

after the stability mark is displayed. Then the
percentage unit is displayed.

Caution
When the sample is too light and it is not possible
to calculate a 100% mass, the scale displays

lo .

and returns to step 3.
Pressing the

ENTER

switch after

lo

is displayed,

the next unit is displayed.

Step 5

Remove all items from the pan.

Advertising