2 preparation – A&D Digital Platform Scales HW-10KGV User Manual

Page 43

Advertising
background image

HW-G Instruction Manual Page 41

HW-G OZIM v1a

14.2 The Complete Calibration Procedure

14.2.1 Gravity Acceleration Correction

Step1

Turn on the display.
Open the rear cover of the display unit.
Inside the display unit, press and hold the

CAL

switch to enter the calibration mode.

Then

Cal 0

is displayed.

Step2

Press the

ZERO

switch to enter the gravity

acceleration correction mode.

Step3

Set your local gravity acceleration using
the following switches.

^

switch Selecting the value of a digit.

<

switch Selecting a digit.

Step4

Press the

ENTER

switch to store the new

value.

Step5

Press the

CAL

switch again. Then the

scale returns to the normal weighing
mode.

14.2.2 Preparation

Step 6 Confirm the environmental conditions as follows:

Maintain a constant temperature and a stable power supply.
Install the scale on a solid floor or bench where there is no draft,
vibration, strong magnetic fields or direct sunlight.
Consider section “4. Caution”.

Step 7 Display normal weighing for at least 30 minutes to warm up the scale.