Disconnect depleted Source A cylinder and connect full cylinder.

Open Source A (left) cylinder valve and manual valve between Source A (left) cylinder and AK M60
crossover manifold. Pressure on Source A (left) gauge should indicate the full cylinder pressure.

AK M60 Model

Source A Depletion

Pressure, psig

AK M6010

85

AK M6015

135

AK M6025

225

Table 3. Approximate Source A Cylinder Depletion Pressure