Troubleshooting

Trouble

Possible cause

Action to be taken

Suction capacity
impaired

Motor speed is reduced Increase motor speed (p. 35)

Check valve closed

Set mode selection switch
for 'power delivery' and press
lower button for blow out
(p. 37)

Filter or check valve
dirty

Clean and/or replace check
valve (p. 41)

Pipette drips

Filter not properly
positioned

Insert filter properly (p. 42)

Adapter or pipette
damaged

Replace adapter or pipette
(p. 41)

Pipette not held
properly

Adapter dirty or dam-
aged

Clean the adapter, allow it to
dry and/or replace it (p. 41)

After connecting
the AC adapter,
the LED-display
of the instrument
flashes or does not
light up

Plug of the battery not
firmly pushed in

Firmly push in the battery plug
(p. 33)

Wrong AC adapter

Only use original AC adapter
(p. 45)

Battery/AC adapter
defective

Replace battery/AC adapter

Instrument defective

Send the instrument in for
repair (p. 47)

Low battery charge Battery problem

Replace battery (p. 40)

Instrument doesn't
work

Internal error
(carry out reset)

Pull out battery plug and press
a pipetting button for 10 sec-
onds. Then firmly push in the
battery plug again (p. 33)

If recommendations in the Troubleshooting section do not solve current
problems, contact your authorized dealer or the manufacturer directly.