Troubleshooting – BrandTech Scientific QuikSip BT-Aspirator User Manual

Page 18

Advertising
background image

- 18 -

Troubleshooting

Problem

Possible cause

Corrective action

Will not aspirate

– Dispensing cartridge not locked
– Dispensing cartridge leaks
– Filling valve leaks or is sticking
– Connection between filling tube and

tube of the cell-culture

leaks

– Tube of cell-culture

blocked or sticking

– Lock dispensing cartridge (

⇒ page 8).

– Mount new dispensing cartridge (

⇒ page 13).

– Clean filling valve (

⇒ page 16) or replace valve and seal.

– Stick adapter firmly onto the filling tube and pull tube firmly

onto the adapter.

– Clean tube or replace.

Piston and dispensing
cartridge difficult to move

– Crystal deposition
– Improper liquid aspirated

– Replace dispensing cartridge (

⇒ page 13).

– Check, if liquid is proper to aspirate.

Liquid behind the piston of
the dispensing cartridge

– Dispensing cartridge leaks

– Replace dispensing cartridge (

⇒ page 13).

Aspirated liquid can't be
delivered into the bottle

– Discharge valve sticking or damaged

– Clean valve (

⇒ page 16) or replace.