Cleaning, Bd c e g – BrandTech Scientific Titrette User Manual

Page 60

Advertising
background image

b

d

c

e

g

60

a) Remove the rear housing and take out the mounting

tool.

b) Pull out the telescoping filling tube and the recirculation

tube (Fig. b).

c) Use the mounting tool to unscrew the filling valve

(Fig. c).

d) If the sealing ring is contaminated or damaged, carefully

remove it with a pair of curved forceps (Fig. d).

e) Clean the filling valve and sealing ring in an ultrasonic

bath, or replace them (Fig. e).

f) Insert the cleaned or new sealing ring, if necessary

(as in Fig. e).

g) Screw in the filling valve first by hand (Fig. g) and then

tighten it with the mounting tool (1/4 turn is sufficient).

6.

Cleaning / replacing the filling valve

Always perform a standard cleaning before dismantling the
instrument!

Cleaning

Dismantling instrument · Component replacement

If the instrument does not fill up, and if some elastic resis-
tance is evident when the piston is rotated upward, then it
is possible that the ball valve is merely stuck.
In this case, loosen the ball valve using light pressure, for
example, with a 20 µl plastic pipette tip (see the figure at
the side).

Note: