Specifications – WIKA DG-10 User Manual

Page 11

Advertising
background image

11

WIKA operating instructions digital pressure gauge, model DG-10

11250313.03 09/2012 GB/D/F/E

GB

3.6 Accuracy
Accuracy at reference conditions

≤ ±0.5 % of span ±1 digit
Including non-linearity, hysteresis, zero offset and end value deviation (corresponds to measured error

per IEC 61298-2).

Zero error for the zero signal

≤ 0.1 % of span (power-up reset)

Adjustability of zero point

≤ 20 % of span (via tare function with model DG-10-E)

Temperature error at 0 ... 60 °C

Mean temperature coefficient of zero point

≤ 0.15 % of span/10 K

Mean temperature coefficient of span

≤ 0.15 % of span/10 K

Long-term stability

≤ 0.2 % of span/year

3. Specifications

Advertising